FCM Flow Cytometry Analysis System Flow Cytometer
CHL-P800
Optical Parameters
Laser source | Fluorescence Channel | Recommended Fluorescent Fuel |
488nm | FL1:525/45nm | FITC, Alexa Flour 488,CFSE, FLUO-3 |
FL2:592/43nm | PE,PI | |
FL3:692/40nm | PE-Cy5.5,PC5,PC5.5,PerCP,PerCP-Cy5.5 | |
FL4:785/62nm | PE-Cy7,PC7 |
Fluid System Parameters
Sample speed adjustment
High Speed | Medium Speed | Low Speed | Sheath Speed |
60uL/min | 35uL/min | 10uL/min | 10mL/min |
Product Performance Features
1,Fluorescence sensitivity:PE≤300MESF,FITC≤100MESF;
2,Fluorescence Linearity:≥0.99;
3,Detection sensitivity of forward angle scattered light :≤400nm;
4,Product resolution (full peak width CV of forward angle scattered light and fluorescence channel):≤2%
5,Resolution of forward angle scattered light and lateral angle scattered light: able to separate red blood cells and platelets in peripheral blood; able to separate the three groups of white blood cells (lymphocytes, monocytes and granulocytes) in peripheral blood.
6,Sample flux: 10000 cells per second;
7,Carryover rate:≤0.1%;
8,Full-matrix fluorescence compensation, automatic compensating calculation;
9,Open sampling platform: applicable to single tube and 96-hole-plate sample-adding system.